{"id":11605,"date":"2023-02-21T08:00:00","date_gmt":"2023-02-21T14:00:00","guid":{"rendered":"https:\/\/blog.misumiusa.com\/?p=11605"},"modified":"2023-03-24T12:24:48","modified_gmt":"2023-03-24T17:24:48","slug":"stable-measurement-tips-for-microbalances-with-ad-weighing-part-3","status":"publish","type":"post","link":"https:\/\/us.misumi-ec.com\/blog\/stable-measurement-tips-for-microbalances-with-ad-weighing-part-3\/","title":{"rendered":"Stable Measurement Tips for Microbalances with A&#038;D Weighing Part 3"},"content":{"rendered":"\n<p>Concluding our series for tips for stable measurements using microbalances with <a href=\"https:\/\/us.misumi-ec.com\/vona2\/maker\/a-and-d\/\" target=\"_blank\" rel=\"noreferrer noopener\">A&amp;D Weighing<\/a>, here are tips 7 thru 12. To catch up on the previous tips, visit the posts below:<\/p>\n\n\n\n<p><a href=\"https:\/\/us.misumi-ec.com\/blog\/stable-measurement-tips-for-microbalances-with-ad-weighing-part-1\/\">Tips 1 \u2013 3<\/a><\/p>\n\n\n\n<p><a href=\"https:\/\/us.misumi-ec.com\/blog\/stable-measurement-tips-for-microbalances-with-ad-weighing-part-2\/\" target=\"_blank\" rel=\"noreferrer noopener\">Tips 4 \u2013 6<\/a><\/p>\n\n\n\n<h2><strong>Tip #7: <\/strong><\/h2>\n\n\n\n<p>Minimize convection flows inside the weighing chamber.<\/p>\n\n\n\n<p><strong>Why?<\/strong><\/p>\n\n\n\n<p>If, for example, you bring in a <a href=\"https:\/\/blog.misumiusa.com\/stable-measurement-tips-for-microbalances-with-ad-weighing-part-1\/\">weighing sample from another location and measure<\/a> it right away, there may be a difference in temperature between the sample and its surroundings.<\/p>\n\n\n\n<p>When the temperature of the sample is higher, a layer of ascending warm air will be generated around it, which will then push the sample upward and make the value appear lighter, as shown by Figure 3. (Conversely, when it is lower, the value will appear heavier.)<\/p>\n\n\n\n<p>As the sample cools down and the convection flow becomes weaker, the display value will gradually increase, making the measurement unstable.<\/p>\n\n\n\n<p>In addition, a convection flow can occur when you open the door to the weighing chamber and replace the air inside. This will bring faint changes in temperature to the chamber and become another factor for drift and poor repeatability. Putting your hand into the weighing chamber will also disturb the temperature in the chamber.<\/p>\n\n\n\n<p><strong>Solution<\/strong><\/p>\n\n\n\n<p>Leave the <a href=\"https:\/\/blog.misumiusa.com\/stable-measurement-tips-for-microbalances-with-ad-weighing-part-1\/\">weighing sample and container near the microbalance,<\/a> or inside the balance on the shelf, for a sufficiently long time for acclimatization to occur before starting measurements.<\/p>\n\n\n\n<p>When placing the weighing sample on the pan, open the door only to a minimum degree, and close it gently but swiftly.<\/p>\n\n\n\n<p>Never grab the sample or container with your hand. Instead, use tweezers that are long enough to reach the pan area without having to put your hand in.<\/p>\n\n\n\n<div class=\"wp-block-image\"><figure class=\"aligncenter size-full\"><a href=\"https:\/\/blog.misumiusa.com\/wp-content\/uploads\/2023\/02\/Tips-for-Stable-Measurement-AD-Weighing-Convection-Flow.jpg\" target=\"_blank\" rel=\"noopener\"><img loading=\"lazy\" width=\"497\" height=\"469\" src=\"https:\/\/blog.misumiusa.com\/wp-content\/uploads\/2023\/02\/Tips-for-Stable-Measurement-AD-Weighing-Convection-Flow.jpg\" alt=\"\" class=\"wp-image-11607\" srcset=\"https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/Tips-for-Stable-Measurement-AD-Weighing-Convection-Flow.jpg 497w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/Tips-for-Stable-Measurement-AD-Weighing-Convection-Flow-300x283.jpg 300w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/Tips-for-Stable-Measurement-AD-Weighing-Convection-Flow-150x142.jpg 150w\" sizes=\"(max-width: 497px) 100vw, 497px\" \/><\/a><figcaption>Convection Flow<\/figcaption><\/figure><\/div>\n\n\n\n<h2><strong>Tip #8<\/strong><\/h2>\n\n\n\n<p>Protect the microbalance from your body heat and breath.<\/p>\n\n\n\n<p><strong>Why?<\/strong><\/p>\n\n\n\n<p>Heat emitted from your body or conveyed by your breath can affect the weighing sensor and cause values to drift. Your breath, if strong, can also create airflows and destabilize measurements.<\/p>\n\n\n\n<p><strong>Solution<\/strong><\/p>\n\n\n\n<p>Only draw as close to the microbalance as is necessary. Take measures to contain your body heat such as wearing a lab coat and mask. Avoid performing measurements immediately after meals, after which your body heat increases. The best measure is to put an external breeze break (draft shield) such as the <a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223013452473\/?HissuCode=AD-1672\">AD-1672<\/a> over the microbalance.<\/p>\n\n\n\n<div class=\"wp-block-image\"><figure class=\"aligncenter size-full is-resized\"><a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223013452473\/?HissuCode=AD-1672\" target=\"_blank\" rel=\"noopener\"><img loading=\"lazy\" src=\"https:\/\/blog.misumiusa.com\/wp-content\/uploads\/2023\/01\/AD-1672-Breeze-Break-AD-Microbalance.jpg\" alt=\"\" class=\"wp-image-11563\" width=\"406\" height=\"402\" srcset=\"https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/01\/AD-1672-Breeze-Break-AD-Microbalance.jpg 812w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/01\/AD-1672-Breeze-Break-AD-Microbalance-300x297.jpg 300w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/01\/AD-1672-Breeze-Break-AD-Microbalance-150x150.jpg 150w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/01\/AD-1672-Breeze-Break-AD-Microbalance-768x760.jpg 768w\" sizes=\"(max-width: 406px) 100vw, 406px\" \/><\/a><figcaption><a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223013452473\/?HissuCode=AD-1672\" target=\"_blank\" rel=\"noreferrer noopener\">AD-1672 Tabletop Breeze Break from A&amp;D Weighing<\/a><\/figcaption><\/figure><\/div>\n\n\n\n<h2><strong>Tip #9<\/strong><\/h2>\n\n\n\n<p>Avoid impacts or tremors to the sensor.<\/p>\n\n\n\n<p><strong>Why?<\/strong><\/p>\n\n\n\n<p>Sensor shocks result in variations in the zero point and reduced repeatability.<\/p>\n\n\n\n<p><strong>Solution<\/strong><\/p>\n\n\n\n<p>Do not open or close the doors roughly causing an impact. Avoid poking the weighing pan with tweezers or dropping the sample on the pan.<\/p>\n\n\n\n<p>Press keys gently, or ideally, use a remote controller. The <a><\/a><a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223011627880\/?HissuCode=AD-8922A\">AD-8922A<\/a> is recommended when the microbalance is on an anti-vibration table, as any vertical force can tilt the balance leading to inaccuracies.<\/p>\n\n\n\n<div class=\"wp-block-image\"><figure class=\"aligncenter size-full\"><a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223011627880\/?HissuCode=AD-8922A\" target=\"_blank\" rel=\"noopener\"><img loading=\"lazy\" width=\"250\" height=\"250\" src=\"https:\/\/blog.misumiusa.com\/wp-content\/uploads\/2023\/02\/AD-Weighing-Remote-Controller.jpg\" alt=\"\" class=\"wp-image-11608\" srcset=\"https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/AD-Weighing-Remote-Controller.jpg 250w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/AD-Weighing-Remote-Controller-150x150.jpg 150w\" sizes=\"(max-width: 250px) 100vw, 250px\" \/><\/a><figcaption><a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223011627880\/?HissuCode=AD-8922A\">AD-8922A A&amp;D Weighing Remote Controller<\/a><\/figcaption><\/figure><\/div>\n\n\n\n<p>It is also advised to perform \u201cpre-loading\u201d \u2013 placing a weight once or twice before starting actual measurements \u2013 to allow the weight sensor a break-in period.<\/p>\n\n\n\n<div class=\"wp-block-image\"><figure class=\"aligncenter size-full is-resized\"><img loading=\"lazy\" src=\"https:\/\/blog.misumiusa.com\/wp-content\/uploads\/2023\/02\/AD-Weighing-Microbalance-Anti-Vibration-Table-Remote-Controller.jpg\" alt=\"\" class=\"wp-image-11609\" width=\"599\" height=\"458\" srcset=\"https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/AD-Weighing-Microbalance-Anti-Vibration-Table-Remote-Controller.jpg 873w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/AD-Weighing-Microbalance-Anti-Vibration-Table-Remote-Controller-300x230.jpg 300w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/AD-Weighing-Microbalance-Anti-Vibration-Table-Remote-Controller-150x115.jpg 150w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/AD-Weighing-Microbalance-Anti-Vibration-Table-Remote-Controller-768x588.jpg 768w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2023\/02\/AD-Weighing-Microbalance-Anti-Vibration-Table-Remote-Controller-80x60.jpg 80w\" sizes=\"(max-width: 599px) 100vw, 599px\" \/><figcaption>Complete Set Up: Microbalance, Anti-Vibration Table, and Remote Controller from A&amp;D Weighing<\/figcaption><\/figure><\/div>\n\n\n\n<h2><strong>Tip #10<\/strong><\/h2>\n\n\n\n<p>Use the internal calibration of the microbalance for a quick check of the environmental conditions at the time<\/p>\n\n\n\n<p>Count the number of seconds required to complete the internal calibration. If it took longer than usual, it is a sign that there is something wrong with the measurement environment and it would be best to delay the measurement until a later time.<\/p>\n\n\n\n<p>Alternatively, with A&amp;D\u2019s BM-20\/22 microbalances, it is possible to run an automatic repeatability test, which automatically measures the internal calibration weight (approx. 20 g) 10 times and calculates repeatability (standard deviation), to inspect the performance under the given environment.<\/p>\n\n\n\n<div class=\"wp-block-image\"><figure class=\"aligncenter size-full\"><a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223011627330\/\" target=\"_blank\" rel=\"noopener\"><img loading=\"lazy\" width=\"250\" height=\"250\" src=\"https:\/\/blog.misumiusa.com\/wp-content\/uploads\/2022\/12\/Micro-Analytical-Balances-Ion-BM-Series.jpg\" alt=\"\" class=\"wp-image-11496\" srcset=\"https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2022\/12\/Micro-Analytical-Balances-Ion-BM-Series.jpg 250w, https:\/\/us.misumi-ec.com\/blog\/wp-content\/uploads\/2022\/12\/Micro-Analytical-Balances-Ion-BM-Series-150x150.jpg 150w\" sizes=\"(max-width: 250px) 100vw, 250px\" \/><\/a><figcaption><a href=\"https:\/\/us.misumi-ec.com\/vona2\/detail\/223011627330\/\" target=\"_blank\" rel=\"noreferrer noopener\">BM Series Microbalance from A&amp;D Weighing<\/a><\/figcaption><\/figure><\/div>\n\n\n\n<h2><strong>Tip #11<\/strong><\/h2>\n\n\n\n<p>Set a consistent time interval for reading the display value<\/p>\n\n\n\n<p>Microbalances can exhibit slightly different behavior depending on the measurement conditions, including the operator at the time. Using a stopwatch, decide how many seconds after placing the sample you find it most comfortable to record the display value. If it was, say, 20 seconds, then always read the values after 20 seconds for all further measurements as well.<\/p>\n\n\n\n<p>Measured this way, microbalances will provide more repeatable, reliable results.<\/p>\n\n\n\n<p>When the sample is powder, adding it to a <a href=\"https:\/\/us.misumi-ec.com\/blog\/stable-measurement-tips-for-microbalances-with-ad-weighing-part-2\/\">weigh boat should be done outside the microbalance<\/a> in order to maintain the same interval as well as to prevent cross-contamination inside the weighing chamber.<\/p>\n\n\n\n<h2><strong>Tip #12<\/strong><\/h2>\n\n\n\n<p>Disable the zero-tracking function when the amount to be weighed is 100\u03bcg or below<\/p>\n\n\n\n<p>When the sample placed is lighter than 100\u03bcg, the measured value increases so slowly that if the zero tracking is active the microbalance may judge it to be a drifting of the zero point. As a result, the display value will remain zero.<\/p>\n\n\n\n<p>Please note that without zero tracking the zero-point drift is visible on the display. It is therefore recommended to re-zero the balance immediately before each measurement.<\/p>\n\n\n\n<p>Thus concludes the series for stable measurements with A&amp;D Weighing, be sure to explore all of their scales and balances through MISUMI <a href=\"https:\/\/us.misumi-ec.com\/vona2\/maker\/a-and-d\/\" target=\"_blank\" rel=\"noreferrer noopener\">here<\/a>.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Concluding our series for tips for stable measurements using microbalances with A&amp;D Weighing, here are tips 7 thru 12. To catch up on the previous tips, visit the posts below: Tips 1 \u2013 3 Tips 4 \u2013 6 Tip #7: Minimize convection flows inside the weighing chamber. Why? If, for example, you bring in a weighing sample from another location [&hellip;]<\/p>\n","protected":false},"author":46,"featured_media":11610,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"om_disable_all_campaigns":false},"categories":[6,1136],"tags":[1245,1309,1319,1318],"aioseo_notices":[],"gutentor_comment":0,"_links":{"self":[{"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/posts\/11605"}],"collection":[{"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/users\/46"}],"replies":[{"embeddable":true,"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/comments?post=11605"}],"version-history":[{"count":5,"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/posts\/11605\/revisions"}],"predecessor-version":[{"id":11719,"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/posts\/11605\/revisions\/11719"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/media\/11610"}],"wp:attachment":[{"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/media?parent=11605"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/categories?post=11605"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/us.misumi-ec.com\/blog\/wp-json\/wp\/v2\/tags?post=11605"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}